图片展示

WIAG-DM online densitometer

    Technical parameters

    Signal mode: Four-wire system (24VDC+4~20mA);

    Output signal: 4-20mA (active output) / 485 modbus RTU communication;

    Power supply: 24VDC (External resistance components cannot be loaded or PLC cannot be connected in series);

    Concentration range: 0-100%(at 20℃). According to the usage conditions, it can be calibrated to a certain range segment.

    Density range: 0-2g/ml. According to the usage conditions, it can be calibrated to a certain range segment.

    Concentration accuracy: 0.5%, resolution: 0.1%, repeatability: 0.2%;

    Density accuracy: 0.003g /ml, resolution: 0.0001, repeatability: 0.0005;

    Explosion-proof grade: Flameproof ExdII BT6

    Applicable pressure: <1Mpa;

    Applicable temperature: 0 to 60℃ (liquid state) Ambient temperature: -40 to 85℃

    Medium viscosity: <5000mpa·s;


  • Parameters

    Technical parameters

    Signal mode: Four-wire system (24VDC+4~20mA);

    Output signal: 4-20mA (active output) / 485 modbus RTU communication;

    Power supply: 24VDC (External resistance components cannot be loaded or PLC cannot be connected in series);

    Concentration range: 0-100%(at 20℃). According to the usage conditions, it can be calibrated to a certain range segment.

    Density range: 0-2g/ml. According to the usage conditions, it can be calibrated to a certain range segment.

    Concentration accuracy: 0.5%, resolution: 0.1%, repeatability: 0.2%;

    Density accuracy: 0.003g /ml, resolution: 0.0001, repeatability: 0.0005;

    Explosion-proof grade: Flameproof ExdII BT6

    Applicable pressure: <1Mpa;

    Applicable temperature: 0 to 60℃ (liquid state) Ambient temperature: -40 to 85℃

    Medium viscosity: <5000mpa·s;



Product Features

It adopts a 4-wire system mode, with 4-20mA output transmission and RS485-Modbus RTU communication

Concentration value: 0-100.0% (20℃) display (1 decimal place); Real-time display of five-digit density values (four decimal places); Baume degree, current value and temperature

Real-time indication enables direct access to the instrument menu on-site for parameter setting and debugging. It is easy to operate and features pure water calibration, fine-tuning of measurement values, and temperature compensation functions

Online real-time measurement can be directly applied to production process control

The titanium alloy material of the parts in contact with liquids can be customized according to requirements

WIAG-DM online densitometer
Long press to look detail
Long by picture save/share

Contact Us

Unit 1804, South Bank Tower, 55 Upper Ground, London, United kingdom, SE1 9EY.

+0044-2084 /323-088

info@wiag-ita.uk

Copyright © WIAG Automation Group All Rights Reserved.

Copyright © WIAG Automation Group 

All Rights Reserved.

添加微信好友,详细了解产品